Chapter 7 - Diagnostics/Troubleshooting
7-17
Section 7-7
Minimum Configuration to Boot/Scan
7-7-1
Booting without using the Ultrasound Hardware
If the low-level ultrasound hardware is failing, the application can not be initialized correctly.
Consequently, proper boot-up of the system is prevented.
By steadily pressing the
H
key on the alphanumeric keyboard during startup, the application can be
started, even if the ultrasound-hardware is broken.
The hardware is not initialized; the application simply does not get ultrasound data into process.
However, one can use SonoView to save images and to gather system information by saving a “Full
Backup”. The remote service capabilities are also present.
Operation:
1.) Power On / Boot Up the system as described in
Section 3-6-2 on page 3-69.
2.) As soon as the start-up process bar indicates
“Initialization of Voluson”
, press the
H
key on the
alphanumeric keyboard.
3.)
Keep the H key pressed during complete startup
(~ 1 minutes). Following message appears:
4.) Click on CONTINUE. The “Probe Selection” menu appears on the monitor.
5.) Press the
FREEZE
key. The application starts, but does not show any Ultrasound information.
6.) Backup all images via SonoView, see:
Section 4-5-6 "Archiving Images" on page 4-37
.
Figure 7-20 Start-up bar: “Initialization of Voluson”
Figure 7-21
with Software 6.x.x and 7.x.x
with Software 8.x.x and higher
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......