GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5-19
5-3-2
Radio Frequency Interface (RBI) Board
The Radio Frequency Interface (RBI) Board combines the following sub-boards and functionalities:
•
FEC (Front End Controller)
•
RFT (Radio Frequency & Tissue) processor
•
IMP (Image Port)
5-3-2-1
Tasks Performed by the RBI Board
•
System Master Clock generator
•
Transmitter Control
•
ATGC Control
•
Receive Focusing Control
•
Analog Test Signal Generator
•
High Voltage Control and supervision
•
Probe Sensing
•
Control of MUX board and RBF boards
NOTE:
Additional Coax Cable need to be used and connected if system is equipped with GPP5/GPP6 and
RBI4 (See connection on GPP and RBI (
Figure 5-6
, above).
Not using the cable shows slight PW-mode disturbances. The Coax Cable prevents it.
5-3-3
RBF Boards (Transmitter/Receiver)
The Transmitter/Receiver has 64 channels and is made up of two identical RBF boards.
Each
RBF board is a complete 32-channel ultrasound Front End that includes the following features
and functions:
•
32 Transmitters
•
32 Receiving channels, T/R switch, Pre-Amp, TGC, ADC
•
32 channel digital Beamformer