Chapter 4 - Functional Checks
4-25
4-4-9
Probe/Connectors Usage
4-4-9-1
Connecting a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) Press the probe connector locking lever
downwards.
Afterwards align the connector with the probe port and carefully push it into place.
4.) Press the connector locking lever
upwards
to the full vertical position to lock it.
5.) Carefully position the probe cord so that it is free to move and is not resting on the floor.
4-4-9-2
Activating the probe
The probe is initialized, the main menu (2D mode) and the ultrasound image appears on the monitor in
write mode (real time display).
4-4-9-3
Deactivating the probe
When deactivating the probe, the probe is automatically placed in standby mode (read mode).
1.) Press the
FREEZE
key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual of
Voluson i / Voluson e for complete cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
4-4-9-4
Disconnecting the probe
Prior to disconnecting a probe freeze the image. It is unnecessary to switch the unit off.
1.) Press the locking lever
downwards
to unlock it.
2.) Carefully slide the probe and connector away from the probe port and around the right side of
the keyboard.
4-4-9-5
Connecting a probe to the Dock Cart / Voluson Station
The Probe Box enables the connection of up to 3 probes to the Voluson i / Voluson e.
1.) Push the probe connector into the probe socket of the Voluson i / Voluson e.
2.) Lift the locking lever upwards to lock the probe connector.
Press the
PROBE
key on the alphanumeric keyboard to activate the “Probe Select” menu.
Select the appropriate probe using the
TRACKBALL
and the
TRACKBALL KEYS
.
Upon selection of an “Application”, the programmed user presets appear.
Clicking a “Setting” field causes loading of the preset.
CAUTION
!! CAUTION:
If a probe is disconnected while running (write mode) a software error may occur.
In this case switch the unit OFF (perform a reset).
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......