Chapter 5 - Components and Functions (Theory)
5-59
5-14-2
Memory Check in LINUX
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see:
Figure 5-32 on page 5-57
), press the
PG DN
[
↓
Arrow down] key on the keyboard until the
Memtest
item is highlighted, then press
ENTER
.
The PC Memory Test will start automatically and will take about 2.5 hours.
If there are errors they will be listed.
NOTICE
!! NOTICE:
After 3 sec. without pressing any key, the system will boot-up in windows.
If you missed step 1, retry again with
CTRL
+
ALT
+
DEL
.
Figure 5-34 Memory check in LINUX
NOTICE
!! NOTICE:
After one cycle (~ 2 hours) the memory check starts again. To interrupt the test, press the
ESC
key.
If you don’t interrupt the memory test by pressing the
ESC
key, it will perform never ending cycles of
memory checks.
elapsed test time
list of errors