GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-55
8-19-11
Replacement of the Probe Holder for Endocavity probes
8-19-11-1
Manpower
One person, 5 minutes
8-19-11-2
Tools
Phillips screwdriver 1
8-19-11-3
Preparations
none
8-19-11-4
Probe Holder- Removal Procedure
1.) Screw off the 2 screws and carefully remove the Probe Holder.
8-19-11-5
Probe Holder- Installation Procedure
1.) Place the new Probe Holder and fasten it with the screws as shown in
Figure 8-75 on page 8-55
.
Figure 8-75 screw off the screws
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......