7-14
Section 7-6 - How to use the Auto Tester program
Section 7-6
How to use the Auto Tester program
NOTE:
Startup the system. Before starting an exam Auto Tester must be turned on.
Immediately after occurrence of the issue you need to stop the Auto Tester and save the results!
In case of a system shut down, the Auto Tester stops automatically and the results will not be saved!
1.) Press the
UTILITIES
key and select the SYSTEM from the menu area on the left side of the screen.
2.) Select the
SERVICE
page on the screen. The “password window” appears automatically.
3.) Enter the password
SHE
and click the ACCEPT button to display the Service Tools window.
4.) Activate the “Auto Tester” program by clicking the START button.
The following message box appears.
5.) Click OK.
6.) Press the
ALT GR
key on the alphanumeric keyboard.
Figure 7-13 Service Tools window
Figure 7-14 Message Box
with Software 6.x.x and 7.x.x
with Software 8.x.x
Start Auto Tester
Start Auto Tester
Alt Gr
key on the
right lower corner
of the keyboard