3-71
3-6-2-1
During a normal boot, you may observe
A.) The ventilation fans starts on full speed, but slow down after a few seconds (listen to fan noise)
B.) Power is distributed to Peripherals, Operator Panel (control panel), Monitor, Front-End and
Back-End Processor.
C.) The Back-End Processor and rest of the scanner starts with the sequence listed in following steps:
1.) The “Boot Screen” is displayed. (
Voluson
is highlighted,
Figure 3-88
below).
2.) Back-End Processor is turned ON and starts to load the software.
3.) The Start Screen (Voluson i) is displayed on the monitor.
4.) Start-up process bars indicating software loading procedures, are displayed on the monitor, as
shown in
Figure 3-89
below.
5.) The software initiates and sets up the Front-End electronics and the rest of the scanner.
6.) The Keyboard backlight is lit.
7.) As soon as the software has been loaded, the 2D screen is displayed on the monitor.
Figure 3-88 Boot screen
Figure 3-89 GE wallpaper with progress bar
with Software 8.x.x and higher
with Software 6.x.x and 7.x.x