Chapter 7 - Diagnostics/Troubleshooting
7-13
7-5-3
CSD: Configuration
Figure 7-11
, above shows the [Device Information] and [Advanced Configuration] tab in the Questra
Agent Configuration menu. This pages normally do not have to be modified.
7-5-3-1
To configure Service Platform
1.) Ask the hospital’s network Administrator for the Proxy Server Address and the Proxy Server Port.
2.) Select the
Configuration
page, then double-click “Questra Agent Configuration”.
3.) Select the [Proxy_Configuration] tab and enter Proxy Server Address and Proxy Server Port.
4.) Click the SAVE button and then close the service page.
Figure 7-11 CSD - Configuration: Device Information / Advanced Configuration
NOTICE
!! NOTICE:
If a Proxy server is available, the
information has to be properly entered, otherwise Questra
remote control does not work
. There is no possibility to detect proxy server information automatically.
Figure 7-12 CSD - Configuration: Proxy Configuration
enter Proxy
Server Information
[Save] button