Chapter 8 - Replacement Procedures
8-49
8-19-5
Replacement of the Back Column Cover
8-19-5-1
Manpower
One person, 15 minutes
8-19-5-2
Tools
Phillips screwdriver 1
8-19-5-3
Preparations
1.) Remove the Cover Bottom as described in
Section 8-19-4 on page 8-48
.
8-19-5-4
Back Column Cover - Removal Procedure
1.) Screw off the 4 screws of the white part and carefully remove.
2.) Screw off the 4 screws of the gray part and carefully remove.
8-19-5-5
Back Column Cover - Installation Procedure
1.) Place the new Back Column Cover and fasten it with the screws.
2.) Mount the Cover Bottom as described in
Section 8-19-4-5 on page 8-48
.
Figure 8-67 screw off the screws