Section 5-13 - Service Page
2.) Select the Check and/or choose correct probe name from the pull down menu.
3.) Check and/or enter probes Serial Number in the appropriate [SNO] field.
4.) Select “Today” from the Start date pull down menu.
NOTE:
DO NOT
disconnect the probe during the re-programming process!
Probe ID and coding is assigned to the probe that is connected to the selected port.
5.) Confirm with
CAUTION
DO NOT assign wrong coding (e.g., programming a RAB to a RIC probe).
An inaccurate probe programming will mostly likely damage the probe and/or system.
Figure 5-31 Probe EEPROM screen
check and/or choose
probe form pull down menu
check and/or enter
probes serial number
select “Today” from
pull down menu
Confirm with OK