Chapter 7 - Diagnostics/Troubleshooting
7-5
Shortcut
Description
Ctrl + C
copy text
Alt + D
export system state (Full Backup and Dump files) to
D:\export
Ctrl + H
shortcuts list (see:
Figure 7-2
, above)
H during start
start UISApp in a mode ignoring all Ultrasound hardware
Ctrl + Alt + I
start/stop Presentation Loop
Ctrl + Alt + L
full Backup load
Alt + P
stores screenshot on D:\export
Ctrl + Alt + S
full Backup save
Ctrl + V
paste, insert text from clipboard
Ctrl + X
cut text
Ctrl + Z
undo typing
Ctrl + F8
show password dialog for enabling shortcuts
Alt + F5
create process logger snapshot
Alt + F6
burn process logger snapshot to CD
Alt + F7
start/stop process logger
Shift + F11
start remote socket control
Alt + F11
stop remote socket control
Alt + Backspace
undo typing
Shift + Delete
cut text
Shift + Insert
insert text (paste)
Ctrl + Insert
copy text