GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-143
P R E L I M I N A R Y
8-6-6 ECG Module Assembly Replacement Procedure
4) Turn the ECG Module assembly (with bracket attached) over, ready to place it in the correct location
as shown in
5.) Carefully slide the assembly and fastening bracket sideways to the
right
until it is seated in the
correct position. Make sure the securing bracket holes are properly aligned -
6) Fasten the three screws on the bracket to secure the ECG Module assembly to the cabinet cage
assembly (refer to
).
7) Return the cabinet cage assembly to the system, reconnect the cables (previously disconnected),
close the cabinet cage door and fasten securely with the two latches.
Refer to the
8-6-1-4 "Cabinet Cage Assembly Closing and Installation Procedure" on page 8-114
8) Return the front cover, as described in the
8-2-6-4 "Front Cover Installation Procedure" on page 8-
.
9) Return the left and right side covers, as described in the
8-2-3-4 "Left Side Cover Installation
8-2-2-4 "Right Side Cover Installation Procedure" on page 8-5
10.)Turn ON power to the system.
Figure 8-175 Return the ECG Module Assembly and Bracket
Figure 8-176 Sliding the ECG Module Assembly and Bracket into Position
ECG Cable
USB Cable
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...