Power Shut Down
When switching the unit OFF, the system performs an automatic shutdown sequence.
NOTE:
After turning OFF the system, wait at least 10 seconds before turning it on again. The system may not
be able to boot-up if power is recycled too quickly.
The system can be switched OFF in one of two ways, as follows:
•
Holding down the
On/Off
button for 3-5 seconds will evoke the display of the System Exit dialog
box (see
Figure 4-7 "System Exit Dialog Box" on page 4-7
). From this, you can either select
Full
Shutdown
(click the
Shutdown
button) or choose to place the system in
Standby Mode
(click the
Standby
button). Both these procedures are described in detail below.
•
Holding down the
On/Off
button for more than 8 seconds will cause the system to perform
Emergency Shutdown.
This is not recommended under normal circumstances.
NOTE:
The system will remain in Standby Mode for 90 minutes.
The System - Exit dialog box, used when switching OFF the unit, provides the following choices:
•
Standby
Use this button to select the Standby mode, allowing a shorter reboot time. Most of the system is
powered down, but a certain portion of the unit remains energized. The standby mode allows a
shorter re-boot time when the system is used on a daily basis or moved from one place to another.
•
Logoff
Use this button to log off the current user.
The system remains ON and ready for a new user to log on.
(If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.)
•
Shutdown
Use this button to Shut Down the system. The entire system is shut down. It is recommended to
perform a full shutdown at least once a week.
•
Cancel
Use this button to close Exit Dialog box - Exit dialog box and return to the previous operation.
•
Exit
This button appears only if the application software was activated from
Maintenance
mode. Use
this button to exit from the System Application software and return to the Windows Desktop.
Figure 4-7 System Exit Dialog Box
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...