3-126
Section 3-4 - Preparing for Installation
P R E L I M I N A R Y
Configure the Search Criterias
NOTE:
This configuration is required.
1.) Click the
Search Criterias
button.
2.) Select the criteria
Modality
and click the
Remove
button.
3.) Press the
OK
button.
Configure Search Criteria: Constrain Query by AE Title
NOTE:
This configuration is optional. It should only be performed if the site is directing orders to the
Vivid S5 N or Vivid S6 N that is scheduled to perform the study.
Please contact the Mitra support personnel to verify that this step is needed.
1.) In the
Search Criterias
dialog box, select
00400001 Scheduled Station AE Title
from the
Select
Tag
pull-down menu.
2.) Enter the Vivid S5 N or Vivid S6 N computer name for the Value.
3.) Click the
Add to List
button.
4.) Click the
OK
button.
Configure Search Criteria: Constrain Query by MRN
NOTE:
This configuration should be performed if the site will need to constrain the Worklist queries by
MRN (patient id)
1.) In the
Search Criterias
dialog box, select
00101000 Other Patient IDs
from the
Select Tag
pull-down menu.
2.) Enter a 9 digit number for the Value.
3.) Click the
Add to List
button.
4.) Click the
OK
button.
Assign DICOM Worklist name to Dataflow
1.) Select the Dataflow tab.
2.) From the
Dataflow name
pull-down menu, select the Dataflow which will be used.
3.) Remove the DICOM Worklist service currently used in the Dataflow.
4.) From the
Select Service
pull-down menu, select the
Gateway Modality Worklist
service.
5.) Click the
Add
button.
6.) Ensure that the service is displaying the value
In
for the Dir parameter and that its role is
Primary
.
Test Connectivity with the Vivid HL7 Gateway
The purpose of this test is to verify that the Vivid S5 N or Vivid S6 N has connectivity with the Vivid HL7
Gateway.
1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality
Worklist service.
2.) Expand the Gateway destination device.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...