Section 8-3 - Control Console Components Replacement Procedures
P R E L I M I N A R Y
8-4-11
LEDs and Soft Menu Board Replacement (Old-type OP)
The procedures described in this section are applicable to any Vivid S5 N / Vivid S6 N system with the
old-type Operator Panel
8-4-11-1
Tools
Use the appropriate flat screwdriver and T10 Torx screwdriver (T8 may also be used), as indicated in
the LEDs and Soft Menu Board replacement procedure.
8-4-11-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
.
8-4-11-3
LEDs and Soft Menu Board Removal Procedure
1.) Remove the Keyboard Assembly, as described in the
“Keyboard Assembly Removal Procedure” on
2.) Remove all the TGC Keycaps, as described in
“Alphanumeric Keycap Removal Procedure” on
3.) Carefully place the Keyboard Assembly upside-down on a flat, clean, stable surface.
4.) Remove the Keyboard Bottom Cover, as described in the
“Keyboard Bottom Cover Removal
5.) Disconnect the two flat cable connectors (one connected to the LED board and one to the Soft
Menu), as indicated (labelled
C
. (Leave the flat cable in position).
6.) Loosen and remove the four star-shape screws that fasten each of the LEDs and Sensor boards in
position, as shown in
NOTICE
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
“Battery Removal Procedure” on page 8-116
.
Figure 8-102 LEDs and Soft Menu Board Removal
LEDs and Sensor Board
LEDs and Sensor Board
Soft Menu Board
Loosen GND Cable
Loosen GND Cable
C
Note:
= cable to be disconnected
C
C
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...