GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
7-86
Section 7-4 - Performing Front End (FE) Diagnostics
P R E L I M I N A R Y
7-4-6 Probe Select Board Diagnostic Tests
7) When finished reviewing the first report, trackball to the
Exit
button and press
Set
to close the View
Report window.
8) Select the second report (
HPF600
); the report is highlighted in the
Reports
list, and the
corresponding results are shown in the Report window below, as shown in
Log - ACW HPF Test - 600" on page 7-86
9) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
10) After reviewing the test results, proceed as follows:
•
Re-boot the system (this is obligatory).
•
Perform calibration, as described in
“DC Offset Calibration” on page 7-27
.
•
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
•
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function (Theory)
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
11.)Trackball to the
Exit
button and press
Set
to close the
View Test Log
dialog box.
12.)Press the <
Esc
> button on the keyboard to close the Diagnostic
Test Window
.
Figure 7-55 View Test Log - ACW HPF Test - 600
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...