Section 4-2 - General Procedures
P R E L I M I N A R Y
NOTE:
In the next step, it is necessary to screw the plastic cable clamp (into the hole provided in the
support column - centrally located at the rear of the system, see
Clamp and Securing Screw" on page 4-4
).
Then, to route the Mains Power cable upwards and through the clamp, and secure the clamp
in position to prevent the cable from dangling down and becoming caught in the wheels when
moving the scanner (see
Figure 4-3 "Mains Power Cable Secured with Plastic Cable Clamp"
).
6.) Proceed as follows:
a.) Loosen the screw to open the plastic cable clamp.
b.) Route the cable
upwards
and through the clamp.
c.) Screw the clamp into position on the support column, carefully tightening the screw sufficiently
to secure the cable firmly in the clamp.
Take care not to overtighten the screw, or damage the cable!
7.) Connect the other end (male plug) of the Mains Power Cable to a hospital-grade mains power outlet
with the proper rated voltage.
8.) Turn ON the circuit breaker at the rear of the system.
The unit is now ready for Power ON/Boot Up.
.
Figure 4-2 Plastic Cable Clamp and Securing Screw
.
Figure 4-3 Mains Power Cable Secured with Plastic Cable Clamp
Screw hole for Securing Clamp
Plastic Cable Clamp
and Screw
Plastic Cable Clamp
Mains Power Cable
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...