GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
Chapter 10 - Care and Maintenance
10-7
P R E L I M I N A R Y
10-4-2
Functional Checks
The functional checks take approximately 60 minutes to perform. Refer to the
Vivid S5 N / Vivid S6 N
User Manual
whenever necessary.
NOTE:
Refer also to
, for additional details about the functional checks described
in this section.
10-4-2-1
System Checks
Table 10-4 System Functional Checks
Step
Item (or Mode)
Description
1
B-Mode
Verify basic B-Mode (2D) operation.
Check the basic system controls that affect this mode of operation.
2
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic system controls that affect this mode of operation.
3
Doppler Modes
Verify basic Doppler operation (PW and CW if available).
Check the basic system controls that affect this mode of operation.
4
M-Mode
Verify basic M-Mode operation.
Check the basic system controls that affect this mode of operation.
5
Applicable Software Options
a
a. Some software may be considered as standard, depending upon system model configuration
Verify the basic operation of all optional modes such as Dual Focus, Virtual convex on
linear probes, SRI, ATO/ASO, APRO, CPI, Compound, AMM, LOGIQ View, B-Flow/BFI,
Stress Echo, TVI/TT, TSI, IMT, LVO Contrast, Vasc./Abd. Contrast, OB Package... etc.
Check the basic system controls that affect the operation of each option.
9
Probe Elements
Perform an Element Test on each probe to verify that all the probe elements and system
channels are functional.
10
System Diagnostics
Perform the Automatic Test to verify that all boards are functioning according to
specifications.
11
Xmit/Recv Elements
Use the Visual Channel Utility on the 12L probe to verify that all system
xmit/recv channels are functional.
12
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
13
LCD Display
Verify basic LCD display functions.
Refer to Chapter 3 of the Vivid S5 N or Vivid S6 N User Manual.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...