GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
Chapter 3 - System Setup
3-115
P R E L I M I N A R Y
The Internet Protocol (TCP/IP) Properties dialog box opens.
6.) Activate the
Use the following IP address
check box.
7) Type in the network computer’s
IP address 10.0.0.2
.
Note:
The IP address must be
different
from the one used for the Vivid S5 N or Vivid S6 N system.
8) Type in the
Subnet mask 255.255.255.0
.
Note:
The Subnet mask should be the
same
for both the network computer and the Vivid S5 N or
Vivid S6 N.
9) Click
OK
, close all windows and approve all changes. Return to regular mode.
10) Connect the Vivid S5 N or Vivid S6 N and network computer LAN connection using a cross over
Network cable. (Normally with RED connector covers on both sides, cable not longer than 100
meters).
The Vivid S5 N or Vivid S6 N and the network computer are now ready for export, import of files.
Note:
You can check the connection by pinging to the Vivid S5 N or Vivid S6 N IP, as follows:
a.) Select <
Start
>, <
Run
>
b.) Type
CMD
, then press <
Enter
>
c.) Type
Ping
and
IP 10.0.0.1
for Vivid S5 N or Vivid S6 N then press <
Enter
>.
If return timing is OK, you have an operating network connection.
3-7-3-8
Configuring the Vivid S5 N or Vivid S6 N and Network PC for a LAN Connection
NOTE:
The unit factory default is DHCP or automatic IP Address assignment. If a static IP Address is required,
use the procedure below.
NOTE:
The following steps are performed on the Vivid S5 N or Vivid S6 N scanner.
The instruction Select = trackball to the specific item and press the Set button on the control panel.
1.) On the control panel, press the
Config
button.
2.) From the System Configuration window, click the
Connectivity
tab at the bottom of the window.
The Connectivity dialog box opens.
3.) Select the
TCPIP Tab.
Figure 3-134 Internet Protocol (TCP/IP) Properties
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...