Section 4-3 - Functional Checks
P R E L I M I N A R Y
ACW Delay Range Test
1
Trackball to
ACW Delay Range Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the ACW Delay Range Test sequence is
complete, the View Test Log window opens
automatically, listing the test performed and showing
whether it passed or failed - see the examples in
on page 7-72 through
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
ACW Pre-Mixer Gain Test
1
Trackball to
ACW Pre-Mixer Gain Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the ACW Pre-Mixer Gain Test sequence is
complete, the View Test Log window opens
automatically, listing the test performed and showing
whether it passed or failed - see the example in
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
ACW BPF Test
1
Trackball to
ACW BPF Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
Table 4-12 Front End Tests (cont’d)
Step
Task
Expected Result(s)
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...