Chapter 7 - Diagnostics/Troubleshooting
7-49
P R E L I M I N A R Y
NOTE:
For this test, there are two reports (MLA0 and MLA1) which is listed in Reports on the right. By
default, the MLA0 report is displayed in the Reports window below, showing the results for both
the TR32A and TR32B boards (
Figure 7-23 "View Test Log - TR32 BF Digital Test (MLA0)" on
9.) To view the MLA1 test results in the Report window, select MLA1 test results under Reports.
The MLA1 report is displayed in the Reports window below:
10) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
11) After reviewing the test results, proceed as follows:
•
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
•
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in
“DC Offset Calibration” on page 7-27
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function (Theory)
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
12.)Trackball to the
Exit
button and press
Set
to close the
View Test Log
dialog box.
13.)Press the <
Esc
> button on the keyboard to close the Diagnostic
Test Window
.
Figure 7-24 View Test Log - TR32 BF Digital Test (MLA1)
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...