GE
P
ART
N
UMBER
FN091065, R
EVISION
2
VS5 N
AND
VS6 N
S
ERVICE
M
ANUAL
4-28
Section 4-3 - Functional Checks
P R E L I M I N A R Y
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the ADC Test sequence is complete, the View
Test Log window opens automatically, listing the test
performed and showing whether it passed or failed -
see the examples in
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
RX TEST
1
Trackball to
RX Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
5
Trackball to the
Start
button and press
Set
The test commences.
When the RX Test sequence is complete, the View
Test Log window opens automatically, listing the test
performed and showing whether it passed or failed -
see the examples in
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
ACW TESTS
.
ACW IQ Symmetry Test
1
Trackball to
ACW IQ Symmetry Test
and press
Set
.
2
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
3
Under the
Special Instructions
area, activate the
Loops
checkbox and select the required number of
Loops
from
the combo box.
4
Adjacent to
On Fail,
select either
Stop
or
Continue
, as
required.
Table 4-12 Front End Tests (cont’d)
Step
Task
Expected Result(s)
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...