Chapter 7 - Diagnostics/Troubleshooting
7-105
P R E L I M I N A R Y
7-5-1 Accessing the Monitoring Diagnostic Test Options
•
If the test fails, repeat the Fan Speed Test.
•
If errors were identified but the results were marginal, perform calibration, as described in
“DC Offset Calibration” on page 7-27
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function (Theory)
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
10.)Trackball to the
Exit
button and press
Set
to close the
View Test Log
dialog box.
11.)Press the <
Esc
> button on the keyboard to close the Diagnostic
Test Window
.
7-5-1-5
Fans Speed Setting Test
NOTE:
For a description of this test, refer to the information in
“Fans Speed Setting Test” on page 7-15
1.) In the
Diagnostic
Test
window, trackball to
Fans Speed Setting
in the list of
Monitoring
tests
available
for selection from the Diagnostic Test Tree (see
Figure 7-71 "Fans Speed Setting Test" on
) and press
Set
. The Fans Speed Test name is displayed under
Execute;
the relevant
system components to be tested are highlighted light blue in the Schematic View (AC/DC, Battery,
PS), as shown below:
NOTICE
In order to access an individual board test, it is necessary to insert the service dongle from system boot-
up and enter the Service Password.
Figure 7-71 Fans Speed Setting Test
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...