Chapter 3 - System Setup
3-65
P R E L I M I N A R Y
One must use the additional power source to activate the Digital Color Printer. However, this must be
powered from the scanner itself, via one of the isolated peripheral power outlets provided on the rear
connectors panel.
The Digital Color Printer can be connected once the system is powered ON, or after shutdown. All
software drivers for the Digital Color Printer are pre-configured and installed designated to be used with
the specific Digital Color Printer supplied by GE HEALTHCARE.
3-5-1-7-1
Connecting the Color Video Printer
1.) Place the Digital Color Printer on a stable surface close to the Vivid S5 N or Vivid S6 N scanner.
2.) Using the power cable (provided with the printer), connect one end of the power cable to the power
IN socket at rear of the printer, and the other end to the printer peripherals outlet at the rear of the
scanner.
3.) Connect the USB cable between the USB port at the rear of the printer and either of the USB ports
(
upper
or
lower
) on the scanner rear connectors panel.
4.) Proceed to perform the following functionality tests:
-
‘Grounding Continuity” on page 10-23
-
‘Chassis Current Leakage Test” on page 10-24
-
‘Color Printer Test” on page 4-33
NOTE:
After connecting the printer to the Vivid S5 N/ Vivid S6 N system, proceed to the following procedures,
as required:
•
‘Configuring Printing Buttons P1 and P2” on page 3-27
•
‘Setting Printing Orientation (Portrait or Landscape) and Paper Size” on page 3-33
WARNING
WARNING
DO NOT ATTEMPT TO USE A DIFFERENT TYPE OF DIGITAL COLOR PRINTER (BRAND OR
MODEL) OTHER THAN THE DIGITAL COLOR PRINTER PROVIDED BY GE. THE
ULTRASOUND SYSTEM IS AN EXTREMELY SENSITIVE AND COMPLEX MEDICAL
SYSTEM. ANY UNAUTHORIZED PERIPHERALS MAY CAUSE SYSTEM FAILURE OR
DAMAGE!
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...