Section 8-6 - Lower Section Components Replacement Procedures
P R E L I M I N A R Y
8-6-1-4
Cabinet Cage Assembly Closing and Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the movie clips
previously referenced in the
Cabinet Cage Assembly Opening and Removal Procedure
.
1.) Fold the cabinet cage assembly in the closed position and fasten securely by tightening each of the
two thumbscrews on either side of the cabinet cage, as shown in
.
2.) Carefully lift the cage into position on the system taking care to mount it properly on the securing
3.) Make sure the securing hinge latch is properly closed - refer to
.
4.) Place the end of the securing hook into the hole on the top of the cage door to fasten the door in
5.) Reconnect the VGA Cable to the Back End.
6.) Fasten the two connectors to connect the KB Data cable to the KB outlet on the Back End and to the
ECG USB connector on the one side; on the other side to the bundle of cables connected to the DIB.
7.) Secure the KB Data cable to the battery bracket using a cable tie.
8.) Reconnect the Docking Connector to the Back End and fasten securely using the two knurled
(ridged) screws on the Docking Connector (refer to
9.) Reconnect the LAN connector to the Back End (refer to
.
10.) Reconnect the end of the USB cable and secure in position with the cable clip, as shown in
.
11) Release the securing hook from the top of cage door, then carefully swing the door to the closed
position - see
. See also note below and
.
NOTICE
When performing the next two steps, make sure the LAN and USB cables are placed
over
the thick
cable - i.e.
not underneath
(see
), otherwise the cage door will not close properly.
Figure 8-142 Cables Re-connected
Cable Clip
Securing
USB Cable
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...