4-3
Introduction
Warning
To prevent possible electrical shock, fire, or personal injury, do
not make connections to the output terminals when voltage is
present. Standby mode is not sufficient to prevent shock
hazard because
could be pushed accidently. Push
and make sure that the Calibrator is in standby before you
make connections to the output terminals.
This chapter gives instructions to operate the Calibrator from the front panel. For a
description of front-panel controls, displays, and terminals, see Chapter 3, “Features”.
Turn on the Calibrator
When the Calibrator is on, the initial display is “Starting Up...” (see below) and it
completes a self-test routine. If a self-test fails, the Control Display shows an error code.
For a description of error codes, see Chapter 7, “Maintenance.”
nn062f.eps
After self-test, the control display shows the reset condition (see below).
nn063f.eps
For a discussion of the softkey selection shown above (auto/locked), see “Auto Range
Versus Locked Range” in this chapter.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...