Remote Commands
Commands
6
6-41
UNCERT?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Uncertainties command) Shows specified uncertainties for the current output. If there are
no specifications for an output, shows zero.
Parameter: 1. (optional) Preferred unit of primary output uncertainty or PCT (default)
2. (optional) Preferred unit of secondary output uncertainty or PCT
(default)
Response:
1. (float) 90-day specified uncertainty of primary unit
2. (float) 1-year specified uncertainty of primary output
3. (character) Unit of primary output uncertainty
4. (float) 90-day specified uncertainty of secondary unit
5. (float) 1-year specified uncertainty of secondary output
6. (character) Unit of secondary output uncertainty.
Example:
UNCERT?
shows
6.120E-01,6.150E-01,PCT,9.50E-02,
1.150E-01,PCT
UUT_FLUSH
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Flush UUT Receive Buffer command) Flushes the UUT receive buffer for data received
from the UUT over the Calibrator rear panel SERIAL 2 TO UUT serial port. The
command can be sent over GPIB or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Parameter: (None)
Example:
UUT_FLUSH
Flush the Calibrator receive data buffer for the UUT.
UUT_RECV?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Receive Data query) Shows data from the UUT in IEEE-488.2 Standard format
over the Calibrator rear panel SERIAL 2 TO UUT serial port. The command can be sent
over GPIB or RS-232 ports, but applies to SERIAL 2 TO UUT serial port operation.
Response:
<data> (binary block data in definite length format from UUT)
Example:
UUT_RECV?
shows
#211+1.0
Shows (for example) a measurement from the UUT. The format is
#2
(two numbers
follow)
11
(characters follow)
+1.0
(11 characters).
UUT_RECVB?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Receive Binary Data query) Shows binary data as integers from the UUT serial
port. Use the UUT_RECV? command instead if receiving ASCII data.
Parameter: (Optional) Maximum number of integers per line
Response:
(Indefinite ASCII) Comma separated integers as follows:
1. (integer) Number of data bytes shown that do not include the count
2. (integer) Data from the UUT serial port as series of comma separated
integers
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...