5502A
Operators Manual
10-4
SC300 Oscilloscope Calibration Option Specifications
These specifications apply only to the SC300 Oscilloscope Calibration Option. General
specifications that apply to the 5502A Calibrator can be found in Chapter 1. The
specifications are correct given the 5502A is operated with the conditions specified in
Chapter 1, and has completed a warm-up period of no less than two times the length of
time the Calibrator was powered off, to a maximum of 30 minutes.
Volt Function Specifications
Volt Function
DC Signal
Square Wave Signal
Into 50
Ω
Into 1 M
Ω
Into
50
Ω
Into 1 M
Ω
Amplitude Characteristics
Range
0 V to
±
2.2 V
0 V to
±
33 V
1.8 mV to 2.2 V p-p
1.8 mV to 105 V
p-p
[1]
Resolution
<
100 V: 4 digits or 10
μ
V, whichever is greater
≥
100 V: 5 digits
Adjustment Range
Continuous
[1]
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(0.25% of output
+
100
μ
V)
[2] [3]
Sequence
1-2-5 (for example 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range
10 Hz to 10 kHz
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(25 ppm of setting
+
15 mHz)
Typical Aberration
within
20
μ
s from leading edge
< (2% of output
+
100
μ
V)
[1]
The square wave signal into 1 M
Ω
is a positive square wave from 1.8 mV to 55 V p-p. From 95 V to 105 V, its output is a
square wave-like signal that alternates between the negative peak and the positive peak, with the centerline at
−
10 V. Signals
between 55 V and 95 V p-p are not available.
[2]
The uncertainty for 50
Ω
loads does not include the input impedance uncertainty of the oscilloscope. Square wave signals
below 4.5 mV p-p have an uncertainty of
±
(0.25% of output
+
200 µV).
[3]
Signals from 95 to 105 V p-p have an uncertainty of 0.5% of output in the frequency range 100 Hz to 1 kHz. Typical
uncertainty is 1.5% of output for 95 V to 105 V p-p signals in the frequency range 10 Hz to 100 Hz, and 0.5% of output in the
frequency range 1 kHz to 10 kHz.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...