Front-Panel Operation
Adjust the Phase
4
4-47
4.
Push the softkey HARMNIC to record the necessary harmonic (1 to 50), with a
maximum frequency output of 10 kHz. For example, record the 7th harmonic (see
below). When the control display shows the necessary value, push
.
nn109f.eps
5.
Push
one or more times to return to previous menus.
Adjust the Phase
When in the dual ac voltage and ac power output modes, you can set the Calibrator to
source two signals with adjustable phase difference. All phase adjustments move the
AUX waveform in relation to the NORMAL waveform. Phase shift adjustments are
recorded as degrees (0 to
±
180.00) or as a power factor (PF). A leading or positive phase
shift causes the AUX waveform to lead the NORMAL waveform. A lagging or negative
phase shift causes the AUX waveform to lag the NORMAL waveform.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...