SC600 Oscilloscope Calibration Option
Calibrate the Time Base of an Oscilloscope
9
9-21
gl010i.bmp
6.
Push
to remove the input signal.
7.
Do this procedure again for the remaining channels on your oscilloscope.
Calibrate the Time Base of an Oscilloscope
The horizontal deflection (time base) of an oscilloscope is calibrated with a method
equivalent to the vertical gain calibration. A time marker signal is made from the
Calibrator and the signal’s peaks are aligned to the graticule line divisions on the
oscilloscope.
The Time Marker Function
The Time MARKER function, which is available through the MARKER menu, lets you
calibrate the timing response of your oscilloscope. To access the MARKER menu, push
the softkey below MODE until “marker” shows.
gjh060.eps
You can push the MODE softkey to move through the functions as shown, or you can
push
to go directly to the SCOPE menu.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...