Remote Commands
Commands
6
6-17
FORMAT
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Format command)
Use with extreme care.
Restores the contents of the nonvolatile
memory device to factory defaults. The memory holds calibration constants and setup
parameters. You lose all calibration data permanently. The CALIBRATION switch on
the rear panel of the Calibrator must be set in the ENABLE position or an execution error
occurs, except for
FORMAT SETUP
.
Parameter:
ALL
(replaces the whole contents with factory defaults)
CAL
(replaces all cal constants with factory defaults)
SETUP
(replaces setup parameters with factory defaults)
Example:
FORMAT SETUP
Replace the setup parameters with the default setup values (below). (The
FORMAT ALL
command is the same as
FORMAT CAL
and then
FORMAT SETUP
.) The
FORMAT
SETUP
command also clears the
*PUD
string (see the
*PUD
command) and
SRQSTR
is
set to “SRQ: %02x %02x %04x %04x” (see the
SRQSTR
command) and
SPLSTR
is set
to “SPL: %02x %02x %04x %04x” (see the
SPLSTR
command).
Features
Temperature
Standard
ITS-90
Display Contrast*
level 7,7
Host Connection
GPIB (IEEE-488)
Display Brightness*
level 1,0
GPIB Port Address
4
RTD Power Up
Default Type
pt385
Serial Ports
8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud
Thermocouple
Power Up Default
Type
K
EOL (end of line)
CRLF
Current Limits
±20.5 A
EOF (end of file)
012,000
Voltage Limits
±1020 V
Remote I/F
term
Remote Commands
SRQSTR
SRQ: %02x %02x %04x
%04x
*PUD string
cleared
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
Defaults
dBm Impedance
600
Ω
FUNC?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Function query) Shows the current output, measurement, or calibration function. See the
response below for output and measurement modes.
Responses:
DCV
(dc volts function)
ACV
(ac volts function)
DCI
(dc current function)
ACI
(ac current function)
RES
(ohms
function)
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...