Remote Commands
Commands
6
6-31
A reset gets the subsequent commands and values:
Command Value Command Value
CUR_POST AUX
RTD_TYPE
<RTD_TYPE_D
value>
DBMZ <DBMZ_D
value>
SCOPE OFF
DC_OFFSET 0V
STBY
(No
output)
DUTY 50PCT
TC_OFFSET
0
CEL
EARTH OPEN
TC_OTCD ON
HARMONIC 1,
PRI
TC_REF
INT
LCOMP OFF
TC_TYPE <TC_TYPE_D
value>
LOWS TIED
TRIG
OFF
OUT 0V,0HZ
Command Value Command Value
OUT_IMP Z1M
WAVE
NONE,NONE
PHASE 0DEG
ZCOMP
OFF
RANGELCK OFF
ZERO_MEAS OFF
TSENS_TYPE TC
Changes made to the setup menus that are not kept in memory are discarded on reset.
Response: (None)
Example: *RST
Put the Calibrator in a reset condition, evoking the commands and values shown above.
RTD_TYPE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Resistance Temperature Detector Type command) Sets the Resistance Temperature
Detector (RTD) sensor type.
Before you use
RTD_TYPE
, select
RTD
with the
TSENS_TYPE
command. After you use
RTD_TYPE
, select the output temperature with the OUT command. Changes in
temperature sensors changes the output to 0
°
C. When set, the Calibrator keeps the RTD
type until power off or reset.
Parameters:
PT385
(100-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_200
(200-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_500
(500-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_1000
(1000-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT3926
(100-ohm RTD, curve
α
=0.003926 ohms/ohm/
°
C)
PT3916
(100-ohm RTD, curve
α
=0.003916 ohms/ohm/
°
C)
CU10
(10-ohm RTD, empirical curve)
NI120
(120-ohm RTD, empirical curve)
Example:
RTD_TYPE PT3926
Set the RTD type to a 100-ohm type, with the pt3926 curve (
α
=0.003926 ohms/ohm/
°
C).
The resistance of 100 ohms refers to the ice point property, (the resistance of the RTD at
0
°
C (32
°
F)).
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...