10-16
4.
Adjust the time base on your oscilloscope to the fastest position available (20.0 or
50.0 ns/div), see below.
P
u
lse a
b
errations
gl007i.eps
5.
Make sure that your oscilloscope shows the correct rise time and pulse aberration
properties.
6.
To remove the input signal, push
.
The Leveled Sine Wave Function
The Leveled Sine Wave (Levsine) function uses a leveled sine wave, whose amplitude
stays relatively constant over a range of frequencies, to verify the bandwidth of the
oscilloscope. When you verify your oscilloscope, you change the frequency of the wave
until the amplitude shown on the oscilloscope drops 30%, which is the amplitude that
refers to the -3 dB point.
To access the Levsine menu, push the softkey below MODE until “levsine” shows, see
below.
Note
Make sure there is no cable connected to TRIG while you use the Levsine function.
gl029i.eps
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...