5502A
Operators Manual
5-14
The characters REMOTE must have shown on the terminal screen as they were
recorded. If they did not show on the screen, but the Control Display changed to
REMOTE CONTROL, then see step 4 of the “RS-232 Host Port Setup Procedure”
and change the REMOTE I/F setting from comp to term.
If nonsense characters showed on the screen, then there is a mismatch of RS-232
parameters. See step 4 of the “RS-232 Host Port Setup Procedure” procedure for the
correct RS-232 settings and then do this procedure again. Start at Step 5.
If no characters showed on the screen, then see step 3 of the “RS-232 Host Port
Setup Procedure” procedure to make sure that serial was selected for the Host port.
Check that you used the correct RS-232 cable. It must be in a null-modem
configuration where the RX and TX lines are reversed (see Appendix C). Also make
sure that you have connected to the correct COM port on the PC.
8.
Type the command LOCAL and push <Enter>. See that the Calibrator Control
Display changes to the reset condition (below).
nn323f.eps
To experiment with other commands in the command set, see Chapter 6, “Remote
Commands.” When finished, select the Exit command from the File menu to close
the Terminal accessory.
Hint: To keep the communication parameters in Terminal for future operations, first
select Save from the File menu and then give a name, for example, host.trm.
Verify RS-232 Host Port Operation with Visual Basic
Complete the subsequent procedure to verify RS-232 (Host) operation with the
Windows-based programming language Visual Basic. This procedure is if you have
completed Appendix D, “Creating a Visual Basic Test Program” to make the group
RS-232 Test.
Complete the subsequent procedure to verify RS-232 operation with Visual Basic.
1.
Complete the “RS-232 Host Port Setup Procedure” before in this chapter to set up the
Calibrator for RS-232 Host port operation. Note the RS-232 Host port parameters
that you selected in this procedure.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...