Remote Operation
Set up the IEEE-488 Port for Remote Control
5
5-5
RS-232
The SERIAL 1 FROM HOST serial port connects the PC and Calibrator, while
the SERIAL 2 TO UUT serial port is used as a pass-through port for commands from the
PC to UUT with the Calibrator. You can write your own computer programs with the
command set, or use the PC as a terminal and enter individual commands, or you can
purchase optional Fluke Calibration MET/CAL Plus software for RS-232 system
operations. Typical RS-232 remote configurations are shown in Figure 5-2.
After you configure the IEEE-488 or RS-232 port for remote operation, you are prepared
to start command set use. The operation of the command set is given in “Use Commands”
in this chapter. A summary of remote commands is in Chapter 6, “Remote Commands.”
Set up the IEEE-488 Port for Remote Control
The Calibrator is fully programmable for to use on the IEEE Standard 488.1
interface bus. The IEEE-488 interface is also in compliance with supplemental
standard IEEE-488.2, which gives more IEEE-488 features. Devices connected
to the IEEE-488 bus are designated as talkers, listeners, talker/listeners, or
controllers. The Calibrator operates as a talker/listener when it is remotely
controlled by an instrument.
A PC with an IEEE-488 interface controls the Calibrator. Compatible software
for IEEE-488 operation can be purchased from Fluke Calibration with
MET/CAL Plus.
When you use the IEEE-488 remote control interface, there are two restrictions:
1.
Number of Devices
A maximum of 15 devices can be connected in one IEEE-488
bus system. For example, one instrument controller, one Calibrator, and thirteen
UUTs.
2.
Cable Length
The total length of IEEE-488 cables used in one IEEE-488 system is
2 meters multiplied by the number of devices in the system, or 20 meters. Use that
which is less. For example, if 8 devices are connected, the maximum cable length is
2 x 8 = 16 meters. If 15 devices are connected, the maximum cable length is 20
meters.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...