6-30
REMOTE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Remote command) Puts the Calibrator into the remote condition. This command
duplicates the IEEE-488 REN (Remote Enable) message. When in the remote condition,
the Control Display shows the softkey “REMOTE CONTROL Go to Local.” Push this
softkey to change the Calibrator to local operation. If the front panel is locked out, the
Control Display shows the softkey “REMOTE CONTROL LOCAL LOCK OUT.” (See
the
LOCKOUT
command.) To unlock the front panel, use the LOCAL command, or cycle
the Calibrator power switch.
Parameter: (None)
Example:
REMOTE
Put the Calibrator in the remote condition and show this condition on the front panel
Control Display with a softkey REMOTE CONTROL.
RPT_STR
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Report String command) Loads the user report string. The user report string can be read
on the Control Display in local operation, and is shown on calibration reports. The
CALIBRATION switch must be set to ENABLE. (Sequential command.)
Parameter:
String of up to 40 characters
RPT_STR?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Report String query) Shows the user report string. The user report string can be read on
the Control Display in local operation, and is shown on calibration reports. (Sequential
command.)
Parameter: None
Response:
(String) a maximum of 40 characters
*RST
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Reset Instrument command) Resets the Calibrator to the power-up condition. *RST
holds off execution of subsequent commands until the reset operation is complete. This
command is the same as if you push the front panel
key.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...