5502A
Operators Manual
4-60
5502A CALIBRATOR
87
TRUE RMS MULTIMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
mA
A
A
m
mV
V
V
OFF
A
COM
V
mA
m
A
PEAK MI
N
MAX
gvx027.eps
Figure 4-17. Cable Connections to Verify an 80 Series High Amps Function
c.
Set the Calibrator output to 3.5 A at 0 Hz and push
. Make sure the error is
in specification limits.
d.
Do this step again with 0 A at 0 Hz. Make sure the error is in specification limits.
e.
Push
on the Calibrator and push the blue key on the DMM to change to ac
measurements.
f.
Set the Calibrator output to 3.5 A at 60 Hz and push
. Make sure the error is
in specification limits.
g.
Do this step again with the Calibrator set to:
AC Current
Frequency
3.5 A
1.0 kHz
10.0 mA
60 Hz
10.0 mA
1.0 kHz
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...