5502A
Operators Manual
4-30
•
I WAVE
(Current Waveform) Selects the waveform for the current output at the
front panel AUX terminals. See “Waveform Types” in this chapter.
•
“LO”s
(Low Potential Output Terminals) The front panel NORMAL LO and AUX
LO terminals must be tied together either at the UUT or at the Calibrator. When tied
at the UUT, select “open.” The default is tied.
•
PHASE
Sets the phase difference between the NORMAL and AUX outputs. See
“Adjusting the Phase” in this chapter.
Set a Dual DC Voltage Output
Note
Tie the terminals NORMAL LO and AUX LO together at the UUT or at the
Calibrator, with the “LO”s softkey selection “tied.”
The Calibrator makes a dual dc voltage output by sourcing one dc voltage on the
NORMAL outputs and a second on the AUX terminals. Complete the subsequent
procedure to set a dual dc voltage output. If you make an entry error, push
one or
more times to clear the display, then enter the value again.
Caution
To prevent damage to the UUT, makes sure the applied voltage
to the UUT is not larger than the rating of the UUT insulation
and the connection wires.
1.
Push
to clear any output from the Calibrator.
2.
Connect the UUT as in the “Connect the Calibrator to a UUT” section of this chapter.
3.
Set the UUT to measure dual dc voltage on the necessary range.
4.
Push the numeric keys and decimal point key to record the necessary voltage output
at the NORMAL terminals (maximum seven numeric keys). For example, 123.4567.
5.
Push
to select the polarity of the voltage (default is +).
6.
Push a multiplier key, if necessary. For example, push
.
7.
Push
.
8.
The Control Display shows the amplitude of your entry for the NORMAL terminals.
For example, 123.4567 mV (see below).
nn071f.eps
Note
Voltage on the AUX output is a maximum of 3.3 V.
9.
Push the numeric keys and decimal point key to record the necessary voltage output
at the AUX terminals (maximum six numeric keys). For example, 234.567.
10.
Push
to select the polarity of the voltage (default is +).
11.
Push a multiplier key, if necessary. For example, push
.
12.
Push
.
13.
The Control Display shows the amplitude of your entries for the NORMAL terminals
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...