SC300 Oscilloscope Calibration Option
Calibrate the Pulse and Frequency Response on an Oscilloscope
10
10-15
Each option in the Edge menu is shown below.
•
OUTPUT @ SCOPE terminal (50
Ω
)
Shows the location and impedance of the
signal output. If the signal does not show on the oscilloscope, push
. To
disconnect the signal, push
.
You cannot change the output impedance in Edge mode.
•
TRIG
If you use the external trigger, use this key to toggle the trigger off and on.
When on, the indication will show “/1” which shows that the external trigger is at the
same frequency as the edge output.
You can also toggle the trigger off and on with
. The external trigger can be
useful for many digital storage oscilloscopes where triggering on fast rise time
signals cannot be done easily.
•
MODE
Shows you are in Edge mode. Use the softkey to change modes and open the
related menus for the other three oscilloscope calibration modes.
Pulse Response Calibration Procedure for an Oscilloscope
This sample procedure shows how to verify the pulse response of the oscilloscope.
Before you verify your oscilloscope, see the oscilloscope manual for the recommended
calibration settings.
Before you start this procedure, make sure that you use the oscilloscope option in Edge
mode. If you are, the Control Display shows the subsequent menu.
gl028i.eps
Do the subsequent sample procedure to calibrate the pulse response.
1.
Connect the 5502A to Channel 1 on the oscilloscope. Select 50
Ω
impedance or use a
50
Ω
termination at the oscilloscope input. Make sure that the
key is lit. This
shows that the signal is connected.
2.
Change the voltage adjustment for the signal so it agrees with the amplitude value
recommended by your oscilloscope manufacturer to calibrate the edge response. The
default is 25 mV @ 1 MHz.
For example, on a Fluke PM3392A oscilloscope, start with a signal of 1 V @ 1 MHz.
3.
Adjust the scale on your oscilloscope to get a good picture of the edge. For example,
on a Fluke PM3392A oscilloscope with a 1 V @ 1 MHz signal, use 200 mV/div.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...