5502A
Operators Manual
9-6
Pulse Generator Specifications
Pulse Generator Qualities
Positive pulse into 50
Ω
Typical rise/fall times
<2 ns
Available Amplitudes
2.5 V, 1 V, 250 mV, 100 mV, 25 mV, 10 mV
Pulse Width
Range
4 ns to 500 ns
[1]
Uncertainty
[2]
5 % of pulse width
±
2 ns
Pulse Period
Range
22 ms to 200 ns (45.5 Hz to 5 MHz)
Resolution
4 or 5 digits depending upon frequency and width
1-Year Absolute Uncertainty at Cardinal Points, tcal
±
5
°
C
±
2.5 ppm
[1]
Pulse width not to exceed 40 % of period.
[2]
Pulse width uncertainties for periods below 2
μ
s are not specified.
Trigger Signal Specifications (Pulse Function)
Pulse Period
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
22 ms to 200 ns
off/1/10/100
≥
1 V
≤
2 ns
Trigger Signal Specifications (Time Marker Function)
Time Marker Period
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
2 to 9 ns
off/100
≥
1 V
≤
2 ns
10 to 749 ns
off/10/100
≥
1 V
≤
2 ns
750 ns to 34.9 ms
off/1/10/100
≥
1 V
≤
2 ns
35 ms to 5 s
off/1
≥
1 V
≤
2 ns
Trigger Signal Specifications (Edge Function)
Edge Signal
Frequency
Division Ratio
Typical Amplitude into
50
Ω
(p-p)
Typical Rise Time
Typical Lead Time
900 Hz to 11 MHz
off/1
≥
1 V
≤
2 ns
40 ns
Trigger Signal Specifications (Square Wave Voltage Function)
Voltage Function
Frequency
Division Ratio
Typical Amplitude into
50
Ω
(p-p)
Typical Rise Time
Typical Lead Time
10 Hz to 10 kHz
off/1
≥
1 V
≤
2 ns
1
μ
s
Trigger Signal Specifications
Trigger Signal Type
Parameters
Field Formats
Selectable NTSC, SECAM, PAL, PAL-M
Polarity
Selectable inverted or uninverted video
Amplitude into 50
Ω
load
Adjustable 0 to 1.5 V p-p
Ω
, (
±
7 % accuracy)
Line Marker
Selectable Line Video Marker
Oscilloscope Input Resistance Measurement Specifications
Scope Input Selected
50
Ω
1
M
Ω
Measurement Range
40 to
60
Ω
500
k
Ω
to
1
.
5 M
Ω
Uncertainty
0.1 %
0.1 %
Oscilloscope Input Capacitance Measurement Specifications
Scope Input selected
1 M
Ω
Measurement Range
5 to 50 pF
Uncertainty
±
(5 % of input + 0.5 pF)
[1]
[1]
Measurement made within 30 minutes of capacitance zero reference. SC600 option must be selected for at least five minutes
prior to capacitance measurement, including the zero process.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...