Remote Commands
Commands
6
6-13
Response:
<value> signed offset amplitude
Example:
DC_OFFSET? shows
+1.44E-03
Shows 1.44 mV as the value of the applied dc offset. If +0.00 is shown, the dc
offset is zero.
DPF
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Displacement Power Factor command) Sets the displacement power factor (phase
angle) between the Calibrator front panel terminals NORMAL and AUX (for sine waves
output only). The NORMAL terminal output is the phase reference. The phase offset is
expressed as the cosine of the phase offset (0.000 to 1.000) and a LEAD (default) or LAG
term, which tells if the AUX output leads or lags the NORMAL output.
Parameters: <value>
,LEAD
<value>
,LAG
Example:
DPF .123,LEAD
Set the current output on the Calibrator AUX terminals to lead the voltage output on the
NORMAL terminals by 82.93 degrees. (Cosine of 82.93 degrees is 0.123, nominal.)
DPF?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Displacement Power Factor query) Shows the displacement power factor (cosine of the
phase angle) between the Calibrator front panel NORMAL and AUX terminals for sine
wave outputs.
Responses: <value>
,LEAD
<value>
,LAG
Example:
DPF?
shows
5.00E-01,LEAD
Shows a leading power factor of .5 when the current output on the Calibrator AUX
terminals leads the voltage output on the NORMAL terminals by 60 degrees. (Cosine of
60 degrees is 0.5.) 0 shows if power factor does not apply to the output.
DUTY
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Duty Cycle command) Sets the duty cycle of the square wave output. The duty cycle is
the percentage of time the waveform is in the positive part of its cycle (1.00 to 99.00
percent). Duty cycle applies only to single-output square waves.
Parameter:
<value> of duty cycle with optional
PCT
(percent) unit
Example:
DUTY 12.34 PCT
Set the square wave duty cycle to 12.34 %.
DUTY?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Duty Cycle query) Shows the value of the square wave output duty cycle (1.00 to
99.00).
Response:
<value> of duty cycle in percent
Example:
DUTY?
shows
1.234E+01
Shows 12.34 % for the value of the square wave duty cycle.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...