5502A
Operators Manual
1-12
Resistance
Range
[1]
Absolute Uncertainty, tcal
±
5
°
C
±
(% of floor)
[2]
Resolution
(
Ω
)
Allowable Current
[3]
(A)
% of output
Floor (
Ω
) Time and temp since ohms
zero cal
90 Day
1 Year
12 hrs
±
1
°
C 7
days
±
5
°
C
0 to 10.999
Ω
0.009
0.012
0.001
0. 01
0.001
1 mA to 125 mA
11 to 32.999
Ω
0.009
0.012
0.0015
0.015
0.001
1 mA to 125 mA
33 to
109.999
Ω
0.007
0.009
0.0014
0.015
0.001
1 mA to 70 mA
110 to
329.999
Ω
0.007
0.009
0.002
0.02
0.001
1 mA to 40 mA
330 to
1.09999 k
Ω
0.007
0.009
0.002
0.02
0.01
1 mA to 18 mA
1.1 to
3.29999 k
Ω
0.007
0.009
0.02 0.2
0.01
100
μ
A to 5 mA
3.3 to
10.9999 k
Ω
0.007
0.009
0.02 0.1
0.1
100
μ
A to 1.8 mA
11 to
32.9999 k
Ω
0.007
0.009
0.2 1
0.1
10
μ
A to .5 mA
33 to
109.999 k
Ω
0.008
0.011
0.2 1
1
10
μ
A to 0.18 mA
110 to
329.999 k
Ω
0.009
0.012
2 10
1
1
μ
A to 50
μ
A
330 k
Ω
to
1.09999 M
Ω
0.011
0.015
2 10
10
1
μ
A to 18
μ
A
1.1 to
3.29999 M
Ω
0.011
0.015
30 150
10
250 nA to 5
μ
A
3.3 to
10.9999 M
Ω
0.045
0.06
50 250
100
250 nA to 1.8
μ
A
11 to
32.9999 M
Ω
0.075
0.1
2500
2500
100
25 nA to 500 nA
33 to
109.999 M
Ω
0.4
0.5
3000
3000
1000
25 nA to 180 nA
110 to
329.999 M
Ω
0.4
0.5
100000
100000
1000
2.5 nA to 50 nA
330 to
1100.00 M
Ω
1.2
1.5
500000
500000
10000
1 nA to 13 nA
[1]
Continuously variable from 0
Ω
to 1.1 G
Ω
.
[2]
Applies for 4-WIRE compensation only. For 2-WIRE and 2-WIRE COMP, add 5
μ
V per amp of stimulus current to the floor
specification. For example, in 2-WIRE mode, at 1 k
Ω
the floor specification within 12 hours of an ohms zero cal for a measurement
current of 1 mA is: 0.002
Ω
+ 5
μ
V / 1 mA = (0.002 + 0.005)
Ω
= 0.007
Ω
.
[3]
Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by Floor
(new)
= Floor
(old)
x
I
min
/I
actual
. For example, a 50
μ
A stimulus measuring 100
Ω
has a floor specification of: 0.0014
Ω
x 1 mA/50
μ
A = 0.028
Ω
, assuming
an ohms zero calibration within 12 hours.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...