SC600 Oscilloscope Calibration Option
Remote Commands and Queries
9
9-31
SCOPE
(IEEE-488, RS-232, Sequential)
Programs the SC600 oscilloscope calibration option hardware, if installed. The
instrument settings are determined by this command’s parameter. Once in SCOPE mode,
use the OUT command to program new output.
OPER, STBY, *OPC, *OPC?, and *WAI all operate as given in Chapter 6. The state of
the oscilloscope’s output while in SCOPE mode is reflected by the bit in the ISR that is
assigned to SETTLED.
The FUNC? query shows SDCV, SACV, LEVSINE, MARKER, EDGE, and
WAVEGEN for the corresponding oscilloscope modes.
Parameters:
OFF
Turns the oscilloscope hardware off. Programs 0V,0 Hz,
output at the NORMAL terminals, standby.
VOLT
AC and dc voltage mode of the oscilloscope. Programs 20 mV
peak-to-peak, 1 kHz, output at the
SCOPE BNC, output
impedance 1 M
Ω
, standby if from OFF or previously in
standby.
EDGE
Oscilloscope Edge mode. Programs 25 mV peak-to-peak, 1
MHz, output at the SCOPE BNC, standby if from OFF or
previously in standby.
LEVSINE
Oscilloscope-leveled sine mode. Programs 30 mV peak-to-
peak, 50 kHz, output at the SCOPE BNC, standby if from OFF
or previously in standby.
MARKER
Oscilloscope Marker mode. Programs the period to 1 ms,
output at the SCOPE BNC, standby if from OFF or previously
in standby.
WAVEGEN
Oscilloscope Wavegen mode. Programs 20 mV peak-to-peak,
square wave, 1 kHz, no offset, output impedance 1 M
Ω
,
standby if from OFF or previously in standby.
VIDEO
Oscilloscope video trigger mode. Programs 100 % output;
ntsc format, mark line 10, odd field.
PULSE
Oscilloscope pulse mode. Programs 100 nanosecond width
pulses every 1 microsecond.
MEASZ
Oscilloscope impedance measurement mode.
OVERLD
Oscilloscope overload test mode. Programs a 5 V dc output
with a 10 second time limit.
Example:
SCOPE VOLT;
OUT -2V, 0 Hz
(dc voltage, -2 V)
SCOPE VOLT;
OUT 4V, 1 kHz
(ac voltage, 4 V peak-to-
peak, 1 kHz.)
SCOPE EDGE;
OUT 0.5V, 5 kHz
(Edge, 0.5 V peak to peak,
5 kHz.)
SCOPE LEVSINE;
OUT 1V, 20 kHz
(Leveled sine wave, 2 V
peak-to-peak, 20 kHz.)
SCOPE MARKER;
OUT 2 MS
(Marker, period of 2 ms.)
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...