The
NEW REF
(New Reference) key is active when error mode is in operation, and sets the
current output value as a new reference for meter error computation.
The
SETUP
(Setup Menu) key puts the Calibrator into setup mode. The setup menu is shown in
the Control Display. Choose Setup options with the softkeys below the Control Display.
The
RESET
(Reset Calibrator) key aborts the Calibrators current operation and changes it to the
power-up default condition, except when it is used by remote control.
The
CE
(Clear Entry) key clears a partially-completed keypad entry from the Control Display. If
there is a partially-completed entry when
is pushed, the output is unaffected.
The
EDIT FIELD
(Edit Output Display Field) key and its left/right arrow keys provide step
adjustment of the output signals. If any of these keys are pushed or the knob is rotated, a digit on
the Output Display is highlighted and the output increments or decrements as the knob is rotated.
If a digit goes past 0 or 9, the digit to its left or right is carried. An error shows on the Control
Display. The difference between the original (reference) output and the new output is shown.
The
and
keys adjust the magnitude of changes by moving the highlighted digit.
lets
you move from voltage or current to frequency and back. In practice, for voltage and current
outputs, the knob and arrow keys are used to adjust output until the UUT reads correctly. The error
display then shows the UUT deviation from the reference.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...