5502A
Operators Manual
1-26
AC Voltage, DC Offset
Range
[1]
(Normal Channel)
Offset Range
[2]
Max Peak
Signal
1-Year Absolute Uncertainty, tcal
±
5
°
C
[3]
±
(% of dc floor)
Sine Waves (rms)
3.3 to 32.999 mV
0 to 50 mV
80 mV
0.1 + 33
μ
V
33 to 329.999 mV
0 to 500 mV
800 mV
0.1 + 330
μ
V
0.33 to 3.29999 V
0 to 5 V
8 V
0.1 + 3300
μ
V
3.3 to 32.9999 V
0 to 50 V
55 V
0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 to 92.999 mV
0 to 50 mV
80 mV
0.1 + 93
μ
V
93 to 929.999 mV
0 to 500 mV
800 mV
0.1 + 930
μ
V
0.93 to 9.29999 V
0 to 5 V
8 V
0.1 + 9300
μ
V
9.3 to 93.0000 V
0 to 50 V
55 V
0.1 + 93 mV
Square Waves (p-p)
6.6 to 65.999 mV
0 to 50 mV
80 mV
0.1 + 66
μ
V
66 to 659.999 mV
0 to 500 mV
800 mV
0.1 + 660
μ
V
0.66 to 6.59999 V
0 to 5 V
8 V
0.1 + 6600
μ
V
6.6 to 66.0000 V
0 to 50 V
55 V
0.1 + 66 mV
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to
±
50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
minimum outputs in each range.
[3]
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output,
±
1 % of the offset range.
AC Voltage, Square Wave Characteristics
Risetime @
1 kHz
Typical
Settling Time @
1 kHz Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range
Duty Cycle Uncertainty
<1
μ
s
<10
μ
s to 1 % of
final value
<2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
±
(0.02 % of 100 ns), 50 % duty cycle
±
(0.05 % of 100 ns), other duty cycles
from 10 % to 90 %
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz
Aberrations
0.3 % of p-p value, from 10 % to 90 % point
<1 % of p-p value, with amplitude >50 % of range
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...