5502A
Operators Manual
9-4
SC600 Oscilloscope Calibration Option Specifications
These specifications apply only to the SC600 Oscilloscope Calibration Option. General
specifications that apply to the 5502A Calibrator can be found in Chapter 1. The
specifications are correct given the 5502A is operated with the conditions specified in
Chapter 1, and has completed a warm-up period of no less than two times the length of
time the Calibrator was powered off, to a maximum of 30 minutes.
Voltage Function Specifications
Voltage Function
DC Signal
Square Wave Signal
[1]
50
Ω
Load
1 M
Ω
Load
50
Ω
Load
1 M
Ω
Load
Amplitude Qualities
Range
0 to
±
6.599 V
0 to
±
130 V
±
1 mV to
±
6.599 V p-p
±
1 mV to
±
130 V p-p
Resolution
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
Resolution
1
μ
V
10
μ
V
100
μ
V
1 mV
10 mV
Adjustment Range
Continuously adjustable
1-Year Absolute Uncertainty,
tcal
±
5
°
C
±
(0.25 % of output
+ 40
μ
V)
±
0.05 % of output
+ 40
μ
V)
±
(0.25 % of output
+ 40
μ
V)
±
(0.1 % of
40
μ
V)
[2]
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Qualities
Range
10 Hz to 10 kHz
1-Year Absolute Uncertainty,
tcal
±
5
°
C
±
(2.5 ppm of setting)
Typical aberration within 4
μ
s from
50 % of leading/trailing edge
<(0.5 % of 100
μ
V)
[1]
Selectable positive or negative, zero referenced square wave.
[2]
For square wave frequencies above 1 kHz,
±
(0.25 % of 40
μ
V).
Edge Specifications
Edge Qualities into 50
Ω
Load
1-Year Absolute Uncertainty,
tcal
±
5
°
C
Rise Time
≤
300 ps
(+0 ps / -100 ps)
Amplitude Range (p-p)
4.5 mV to 2.75 V
±
(2 % of 200
μ
V)
Resolution 4
digits
Adjustment Range
±
10 % around each sequence value
(indicated below)
Sequence Values
5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80
mV, 100 mV, 200 mV, 250 mV, 300 mV,
500 mV, 600 mV, 1 V, 2.5 V
Frequency Range
[1]
900 Hz to 11 MHz
±
(2.5 ppm of setting)
Typical Jitter, edge to trigger
<5 ps (p-p)
Leading Edge Aberrations
[2]
within 2 ns from 50 % of rising edge
<(3 % of 2 mV)
2 to 5 ns
<(2 % of 2 mV)
5 to 15 ns
<(1 % of 2 mV)
after 15 ns
<(0.5 % of 2 mV)
Typical Duty Cycle
45 % to 55 %
Tunnel Diode Pulse Drive
Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 to 100 V p-p.
[1]
Above 2 MHz rise time specification <350 ps
[2]
All edge aberration measurements made with Tektronix 11801 mainframe with SD26 input module.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...