Remote Operation
Use Commands
5
5-29
Table 5-4. IEEE-488 Interface Messages (Received)
Mnemonic Name
Function
ATN Attention
A control line that, when asserted, tells all instruments on
the bus that the subsequent data bytes are an interface
message. When ATN is low, the subsequent data bytes
are interpreted as device-dependent or common
commands addressed to a specified instrument.
DAC
Data Accepted
Sets the handshake signal line NDAC low.
DAV
Data Valid
Asserts the handshake signal line DAV.
DCL
Device Clear
Clears the input/output buffers.
END End
A message that occurs when the Controller asserts the
EOI signal line before it sends a byte.
GET Group
Execute
Trigger
Trigger a TC measurement and put the indication in the
output buffer.
GTL Go
To
Local
Transfer control of the Calibrator from one of the remote
conditions to one of the local conditions, see Table 5-1.
LLO Local
Lockout
Transfers remote/local control of the Calibrator, see
Table 5-1.
IFC Interface
Clear
A control line that sets the interface to a quiescent
condition.
MLA
My Listen Address
Addresses a specified device on the bus as a listener.
The controller sends MLA automatically when it tells a
device-dependent or common command to a specified
instrument.
MTA
My Talk Address
Addresses a specified device on the bus as a talker. The
controller sends MTA automatically whenever it directs a
device-dependent or common query to a specified
instrument.
REN Remote
Enable
Transfer remote/local control of the Calibrator, see Table
5-1.
RFD
Ready For Data
Sets the handshake signal line NRFD low.
SDC
Selected Device Clear
Does the same thing as DCL, but only if the Calibrator is
currently addressed as a listener.
SPD
Serial Poll Disable
Cancels the effect of a Serial Poll Enable.
SPE
Serial Poll Enable
After the Calibrator receives this message, it will show the
Status Byte the next time it is addressed as a talker over
the GPIB interface.
UNL Unlisten
“Unaddresses” a specified device on the bus as a listener.
The controller sends UNL automatically after the device
has successfully received a device-dependent or common
command.
UNT Untalk
“Unaddresses” a specified device on the bus as a listener.
The controller sends UNL automatically after the device
has successfully received a device-dependent or common
query.
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...