Front-Panel Operation
Use the Setup Menu
4
4-5
•
INSTMT SETUP
(Instrument Setup)
Lets you change the power-up or reset default setting for different instrument
parameters. Many of the same parameters in this menu can be changed during
operation, but the changes you make during operation are volatile. Change them here
to makes them nonvolatile. To set the Calibrator to its factory defaults, use the
Format NV Memory menu in the UTILITY FUNCTNS menu.
•
UTILITY FUNCTNS
(Utility Functions)
Lets you to start self-tests, format the nonvolatile memory (set the factory defaults),
and examine the instrument configuration software versions and user report string.
These features are given in the “Utilities Function Menu” section of this chapter.
Use the Instrument Setup Menu
The softkeys in the instrument setup menu (accessed by pressing INSTMT SETUP
softkey in the Setup Menu) are shown below.
nn065f.eps
The list below gives submenus accessed by each softkey.
•
OTHER SETUP
Opens a menu that lets you toggle the degree reference between the 1968
International Provisional Temperature Standard (IPTS-68) and the 1990 International
Temperature Standard (ITS-90) (factory default). This is also where you set the
clock, and set the power-up and reset defaults for the SC-600 Oscilloscope
Calibration Option Overload test safety timeout function (OVLD T), and shown error
units. You can also configure the Calibrator for best operation with a 50 Hz line
frequency.
•
OUTPUT SETUP
Opens a menu to change the power-up and reset defaults for current and voltage
output limits, default thermocouple and RTD types, set the phase reference, internal
or external phase reference source, and impedance for dBm display.
•
DISPLAY SETUP
Opens submenus to set the intensity and contrast of the Control and Output
Displays.
•
REMOTE SETUP
Lets you change the configuration of the two RS-232 ports, SERIAL 1 FROM HOST
and SERIAL 2 TO UUT, and IEEE-488 General Purpose Interface Bus (GPIB). (See
Chapter 5, “Remote Operation”.)
Содержание 5502A
Страница 4: ......
Страница 12: ...5502A Operators Manual xii ...
Страница 46: ...5502A Operators Manual 2 2 ...
Страница 52: ...5502A Operators Manual 2 8 ...
Страница 54: ...5502A Operators Manual 3 2 ...
Страница 70: ...5502A Operators Manual 3 18 S1 S to T to S1 gvx032 eps Figure 3 4 SETUP Softkey Menu Displays cont ...
Страница 142: ...5502A Operators Manual 4 68 ...
Страница 193: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10 ...
Страница 194: ...5502A Operators Manual 6 2 ...
Страница 240: ...5502A Operators Manual 7 2 ...
Страница 260: ...5502A Operators Manual 7 22 ...
Страница 261: ...8 1 Chapter 8 Accessories Title Page Introduction 8 3 Rack Mount Kit 8 4 IEEE 488 Interface Cable 8 4 5500A LEADS 8 4 ...
Страница 262: ...5502A Operators Manual 8 2 ...
Страница 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 366: ...5502A Operators Manual B 2 ...
Страница 368: ...5502A Operators Manual B 4 ...
Страница 378: ...5502A Operators Manual D 6 ...