Proservo NMS5
98
Hauser
11.2
Alarm History Display
Item
Procedure
Remarks
Matrix Group: DEVICE DATA
Figure. 88:
Device Data
Touch Control
Figure. 89:
LCD (Display)
1. Bring up Static Matrix
GVH=034 *ALARM
CONTACT."
Static Matrix
Figure. 90:
Static Matrix
2. Previous static matrix
records are sequentially
displayed in Static
Matrix screen
GVH=035, starting with
the latest record. Up to
100 alarm records can be
saved. If the number of
records exceeds 100, it is
sequentially
overwritten starting with
the oldest record. The
display includes the year,
month, day, hour, min-
ute, instrument
temperature, and an
error sequential
number, in this order.
For example, 97
3192238 2402 means
that an error occurred at
22:38 on March 19,
1997, when the
instrument temperature
was 24
o
C and this is the
second error since
NMS5 was installed.
MORE
FUNCTION
ALARM CONTACT
0 3 4
MORE
FUNCTION
(Alarm Message)
0 3 5
11 Displacer and Measuring Wire