Proservo NMS5
120
Hauser
0
Matrix group
Function group
Item
Access
code
Short description
Default value
Set
Select
Display
Possible settings,
selection, or display
Index No,
GVH
STATIC MATRIX
(This word is not
shown)
MORE FUNCTION
MATRIX OF
Use (+) and (-) keys, scroll through 9 Dynamic Matrix Group
selections. Press "E" key to select one for access to more data.
CALIBRATION
DEVICE DATA
SERVICE
TEMPERATURE
HART DEVICE (1)
HART DEVICE (2)
ADJ. SENSOR
030
ALARM CONTACT
0
Display alarm message depending on current status.
Display
034
(Calendar)
0
Displays date, time as yymmdd __hh:mm:ss.
Not transferred by Rackbus.
CALIBRATION
Select
Display
Japanese local
time
e.g. 5 410 19:10:41
Year Month Day
HH:MM:SS
033
NO ALARM
035
(Alarm History)
0
Use (+) and (-) keys, scroll through history of alarms.
Up to 99 alarms recorded, then oldest alarm is overwritten
by new data.
Display
DIAGNOSTIC CO
0
Displays current self-diagnostic code
Display
NO ALARM
NO ERROR
036
Alarm message
Alarm history
Error message
(Error History)
0
Use (+) and (-) keys, scroll through history of errors, in format
yymmdd hhmm°° ##. Up to 99 alarms recorded, then oldest
alarm is overwritten by new data.
MPU:
START ACT
Display
Error history
037
RESET ALM. DIAGNO
0
Reset Error History by setting = ON.
OFF
Display
038
ACCESS CODE
0
Set access code to view and change to matrix data. Higher access
codes (except 777) give greater editing privileges.
0
039
Set
50: Operator, 51: technician,
777: required to select EEx i HART line,
0, 50, 51, 530, 777